Instruction Manual EL-FLOW® Select9.17.099F 43
Fluid Set Index
Type
Access
Range
FlowDDE
FLOW-BUS
Modbus
Unsigned char
RW
0…7
24
1/16
0x0030/49
With this parameter, any of the pre-configured fluids (up to 8) can be selected. Each fluid has its specific (configurable)
properties, such as Fluid Name, Capacity, etc.
Default value: 0 (fluid 1).
Fluid Name
Type
Access
Range
FlowDDE
FLOW-BUS
Modbus
Unsigned char[10]
RW B
-
25
1/17
0x8188…0x818C/33161…33165
This parameter contains the name of the current fluid.
Capacity
Type
Access
Range
FlowDDE
FLOW-BUS
Modbus
Float
RW B
1E-10…
1E+10
21
1/13
0x8168…0x8169/33129…33130
This parameter sets the maximum readout/control value (100%) for the current fluid in readout units corresponding to
Capacity Unit.
Capacity Unit
Type
Access
Range
FlowDDE
FLOW-BUS
Modbus
Unsigned char[7]
RW B
see below
129
1/31
0x81F8…0x81FB/33273…33276
Available units:
Mass flow
Normal volume flow
(1.01325 bar(a), 0 °C)
Standard volume flow
(1.01325 bar(a), 20 °C)
Custom volume flow
(Capacity Unit Type Pressure,
Capacity Unit Type
Temperature)
ug/h, ug/min, ug/s,
mg/h, mg/min, mg/s,
g/h, g/min, g/s,
kg/h, kg/min, kg/s
uln/h, uln/min, uln/s,
mln/h, mln/min, mln/s,
ln/h, ln/min, ln/s,
ccn/h, ccn/min, ccn/s,
mm3n/h, mm3n/m,
mm3n/s,
cm3n/h, cm3n/m, cm3n/s,
m3n/h, m3n/min, m3n/s,
scfh, scfm, scfs,
sccm, slm
uls/h, uls/min, uls/s,
mls/h, mls/min, mls/s,
ls/h, ls/min, ls/s,
ccs/h, ccs/min, ccs/s,
mm3s/h, mm3s/m, mm3s/s,
cm3s/h, cm3s/m, cm3s/s,
m3s/h, m3s/min, m3s/s
ul/h, ul/min, ul/s,
ml/h, ml/min, ml/s,
l/h, l/min, l/s,
cc/h, cc/min, cc/s,
mm3/h, mm3/m, mm3/s,
cm3/h, cm3/m, cm3/s,
m3/h, m3/min, m3/s,
cfh, cfm, cfs
Because of the maximum string length (7 characters), some unit names are abbreviated. For instance mm3n/m means
mm
3
n/min.
Capacity Unit Type Temperature
Type
Access
Range
FlowDDE
FLOW-BUS
Modbus
Float
RW B
-273.15…
3.4E+38
245
33/10
0xA150…0xA151/41297…41298
This parameter defines a reference temperature for conversion of the measured mass flow to a volume flow. See also
parameters Capacity Unit and Counter Unit.