5.3 Common issues
No (fieldbus) communication
· Check power supply
· Check cable connection
· Check cable hook-up
Change node address (see Network
configuration)
Reset instrument and/or restart master. If
problem persists, contact Bronkhorst.
· Check power supply
· Check cable connection
· Check cable hook-up
Invalid control mode (instrument
accepts no setpoint)
Check control mode (see Special parameters)
No setpoint given or setpoint too
low
Control valve in Safe State
(normally closed)
Check if control valve is in safe state; solve
cause if necessary (see Valve Safe State)
Inlet pressure or differential
pressure too low
Piping, filters and/or control valve
clogged or blocked
· Flush fluid system with clean, dry air. If
problem persists, contact Bronkhorst.
· For external proportional control valves:
supply 0…15 Vdc and operational inlet
pressure to valve and slowly increase
voltage. If valve does not open, clean parts
and re-adjust valve
Laminar flow element clogged or
blocked
Return equipment to factory
Return equipment to factory
Control behavior unstable
Measurement disturbed by
vibrations
If possible, avoid installation in close proximity
of mechanical vibration
Install pressure regulator or increase buffer
volume between controlling instruments (see
section Piping requirements)
Inlet and/or outlet pressure too
high or too low
Adjust pressure and/or set instrument pressure
in accordance with actual process pressure (e.g.
with FlowTune™)
Wrong process gas selected
Select correct process gas (e.g. with FlowTune™)
Wrong controller settings
Adjust settings (e.g. with FlowPlot)
Return equipment to factory
No flow (sending a setpoint has no
effect)
Check upstream components for obstruction,
e.g.:
· fluid lines
· valves
· filters
Inlet pressure or differential
pressure out of bounds
Set inlet pressure to a value within
specifications