Bronkhors
Instruction Manual EL-FLOW® Select 9.17.099F24
Module status
Pattern
Time
Indication
off
continuous
No power
on, green
continuous
Normal operation mode
blinking,
green
0.5 sec on,
0.5 sec off
Device is in standby mode or configuration is missing, incomplete or incorrect
/
alternating
0.5 sec green,
0.5 sec red
Self test mode
on, red
continuous
Critical error; the instrument needs servicing before it can be used
3.5.2 Multifunctional switch
Some special functions of the instrument can be started manually using the multifunctional switch near the indication
LEDs. These functions are available in analog as well as in digital operation mode.
3.5.2.1 Normal operating functions
· In order to access these functions, press and hold the switch while the instrument is in normal operation mode (green
LED glowing).
· As long as the switch is held, the LEDs show a repeating sequence of patterns, where each pattern indicates a function.
· All indications in this sequence are continuous.
· Each pattern is shown for a number of seconds; in the table below the column Hold time indicates the time frame within
the sequence where the LEDs show the associated pattern.
· To start a function, release the switch when the LEDs show the pattern of the required function.
(green)
(red)
Hold time
Function
off
off
0…1 sec
No action
off
off
1…4 sec
Reset alarm
off
on
4…8 sec
Reset instrument; clear all warnings and error messages and restart the instrument
on
off
8…12 sec
Auto-zero; re-adjust the zero-point of the instrument (flow meters/controllers)
on
on
12…16 sec
Enable FLASH mode for firmware update:
· the instrument shuts down and both LEDs are switched off
· at the next power-up, the instrument will be active again
See Adjusting zero point for background information and instructions on how to adjust the zero point of an instrument.
Never perform a zeroing procedure before having taken notice of the instructions.