Design and Function
16 H155936_3_003
In addition, a sample line holder (shown in the following figure, on top of the magnet), a spare
parts  kit  and  documentation  are  provided.  Refer  to  Connecting  the  Transfer  Line  Holder
[}24] for instructions on how to install it.
Figure4.2: Recommended Laboratory Setup
The glass flow tube with flow tube protector and flow interface are shown in the figure Figure
4.3 [}17].
The  5  mm  NMR  glass  flow  tube  together  with  the  tube  holder,  which  has  the  shape  of  a
spinner, is compatible with any standard Bruker 5 mm NMR probe with insertion depth of 20
mm. The InsightMR glass flow tube volume corresponds approximately to the active volume
of a standard 5 mm tube.
The  transfer  line  is  composed  of  2  sample  lines,  sample  in  and  sample  return,  and  2
concentric  temperature  control  lines  surrounding  the  sample  lines.  Thermostatic  liquid  is
flushed through the temperature control lines to maintain the temperature during the sample
transfer to the NMR and back. Insulating foam around the transfer line prevent temperature
loss, condensation and injury through high temperatures.
The recommended laboratory set up is shown in the figure Figure 4.2 [}16]. The pump and
the thermostat can be purchased together with InsightMR, as options. They are not included
as default.
The reaction mixture is pumped from the vessel, through the pump, the valve (flow interface)
and the sample lines, to the NMR and back. Tubing and standard HPLC fittings are provided
to connect the vessel to the pump, the pump to the valve and the valve back to the reactor.
The temperature control  line  inlets  and  outlets are fixed  to  the  flow  interface, which can  be
secured to a laboratory stand.