Technical Data
H155936_3_003 53
12 Technical Data
NMR System Requirements
• 5 mm probe.
• 300-950 MHz magnet.
• AVIII console if InsightMR software is to be used (compatible with TopSpin 3.5pl2 and
above).
Data Acquisition and Analysis
It  is  highly  recommended  that  InsightMR  software  (refer  to  the  InsightMR  release  letter)  is
used for data acquisition and analysis.
Recommended Temperature Range
The  range  in  which  temperature  loss  was  measured  is  -5  °C  to  +60  °C.  Temperature  loss
using a thermostat type like the Julabo FE25-HE at -5 °C is < 4 °C and at +60 °C is < 2 °C
(remark:  the  loss  is  very  dependent  on  the  flow-power  a  thermostat  can  create,  therefore
these numbers are based on the thermostat type we recommend).
The  unit  can  be  used  at  a  wider  range  of  temperatures  according  to  the  recommended
working temperatures of the wet materials, probe, pump and chiller.
Recommended Pressure
The recommended pressure is < 10 bar.
Wetted Materials for the Sample and Temperature Control Lines
Wetted materials for the sample lines are glass, PTFE (Teflon
®
) and FFKM (Kalrez
®
).
Wetted materials for the temperature control lines are PTFE (Teflon
®
), PEEK, POM (Delrin
®
).
Note:  The  temperature  control  line’s  wet  material  only  come  in  contact  with  the  thermostat
liquid.