Flow Tube Assembly - Insertion Checklist
H155936_3_003 45
8.5 Insert the Flow Tube Assembly
1. Temporarily switch the sample flow off.
2. Inspect the flow tube assembly and the transfer line. Remove any condensation from the
outside of the glass flow tube, the tube holder and the holder body.
3. Hold the flow tube assembly; insert it in the magnet and carefully push it down until the
glass flow tube seats on the probe. To ensure proper insertion, lift it 1 cm up and push it
back down. You must feel a “solid” resistance.
4. Optionally acquire a BESTPROFILE and compare the result to that obtained from a
standard 5 mm sample.
5. Switch the sample flow on.
6. Fix the transfer line to the holder on top of the magnet or use ‘Velcro’ tape to secure it.
7. Block the path where the transfer line is installed to ensure that the transfer line is not a
tripping hazard.
8.6 The System is Ready to be Used
Refer to the InsightMR software quick guide (provided or under InsightMR help menu) for a
quick start.