LYNXEYE XE Detector  User Manual 
DOC-M88-EXX240 V1 – 03.2014  6-5 
ASIC Swapping: Always 0x01 
ASIC Configuration: 
-  0x2C: high count rate, high gain for Cr, Co and Cu rad. 
-  0x2F: high energy resolution, high gain for Cr, Co and Cu rad. (standard for powder diffrac-
tion) 
-  0x24: high count rate, low gain for Mo and Ag rad. 
-  0x27: high energy resolution, low gain for Mo and Ag rad. 
 
5.  Selecting AOUT plug for LYNXEYE XE: Depending on your goniometer type, (Th/Th or Th/2Th) 
you must select the correct AOUT source for the LYNXEYE XE scans: 
 
Figure 6.5: Inner_Circle_Motor for Th/Th-diffractometer  
 
Figure 6.6: Outer_Circle_Motor for Th/2Th-diffractometer 
6.  You now can save and download this configuration and the detector will work but without energy 
discrimination or OpCo recognition. If you insert the Chip ID for your LYNXEYE XE, the Detector 
Optics and its two Detector Optics Mounts, OpCo recognition is working.