LYNXEYE XE Detector User Manual
DOC-M88-EXX240 V1 – 03.2014 7-3
7.2.3 The detector is not Selectable/Active in COMMANDER
If this happens after download, press the RESET button of the detector and wait for reconnect.
Check the LAN connection
Is the detector correct mounted?
Is the OpCo manager running
Is the Chip ID set correctly.
7.2.4 High Background in Scans
Fluorescence from sample: adjust lower discriminator setting.
Detector Bias not set correctly.
One strip became noisy: perform a fixed scan for all strips to identify the noisy one, contact Bruker
AXS.
7.2.5 High Kβ-Part in Scans
Check energy position for diffraction wavelength and compare with actual discriminator settings.
Position might differ due to temperature change of detector.
7.2.6 Alarm LED Lights Red
Reset the detector and wait for reconnection.
7.2.7 The BIAS LED is not Illuminated.
Check the Bias Value in the TOOLS.
7.2.8 The Angular Accuracy Peak Position Data is Larger than
+0.01° of the Expected Value.
Repeat the calibration of the LYNXEYE XE detector.