LYNXEYE XE Detector User Manual
DOC-M88-EXX240 V1 – 03.2014 6-17
Table 6.2: Detector parameters and settings for various target materials
Tube Energy [eV] Mode Gain Peak Pos [V] Energy Resol.[eV] Kβ part DAC Veto Thr.
[V]
Cr 5414 High Res. High 0.141-0.155 <680 <8% 0.040
Cr 5414 High CPS High 0.149-0.162 <1100 --- 0.060
Co 6930 High Res. High 0.180-0.198 <680 <3% 0.040
Co 6930 High CPS High 0.190-0.216 <1100 --- 0.060
Cu 8048 High Res. High 0.209-0.241 <680 <2% 0.040
Cu 8048 High CPS High 0.222-0.252 <1100 --- 0.060
Mo 17444 High Res. Low 0.226-0.261 <740 <1% 0.026
Mo 17444 High CPS Low 0.238-0.274 <1150 --- 0.040
Ag 22106 High Res. Low 0.286-0.332 <780 <0.5% 0.026
Ag 22106 High CPS Low 0.300-0.350 <1150eV --- 0.040
Note
Use a beta filter to determine the correct energy resolution. In High Resolution Mode the
max. count rate in one strip should not exceed 10000 cps and in in High Count Rate mode
100000 cps to avoid significant dead time effects