LYNXEYE XE Detector  User Manual 
DOC-M88-EXX240 V1 – 03.2014  6-17 
Table 6.2: Detector parameters and settings for various target materials 
Tube  Energy [eV]  Mode  Gain  Peak Pos [V]  Energy Resol.[eV]  Kβ part  DAC Veto Thr. 
[V] 
Cr 5414  High Res. High 0.141-0.155 <680  <8%  0.040 
Cr 5414  High CPS High 0.149-0.162 <1100  ---  0.060 
Co 6930  High Res. High 0.180-0.198 <680  <3%  0.040 
Co 6930  High CPS High 0.190-0.216 <1100  ---  0.060 
Cu 8048  High Res. High 0.209-0.241 <680  <2%  0.040 
Cu 8048  High CPS High 0.222-0.252 <1100  ---  0.060 
Mo 17444  High Res. Low 0.226-0.261 <740  <1%  0.026 
Mo 17444  High CPS Low 0.238-0.274 <1150  ---  0.040 
Ag 22106  High Res. Low 0.286-0.332 <780  <0.5% 0.026 
Ag 22106  High CPS Low 0.300-0.350 <1150eV  ---  0.040 
 
 
Note 
Use a beta filter to determine the correct energy resolution. In High Resolution Mode the 
max. count rate in one strip should not exceed 10000 cps and in in High Count Rate mode 
100000 cps to avoid significant dead time effects 
 
 
Figure 6.18: Switch to ‘Application’ and ‘Calibration’