User Manual LYNXEYE XE Detector
6-32 DOC-M88-EXX240 V1 – 03.2014
Figure 6.31: Position check of the ‘Resolution’ and ‘Offset’ determination
The detector is now completely adjusted concerning energy calibration (for one mode: Cu radiation
and high energy resolution) and spatial calibration.
At this point, the Configuration and the State file should be saved (e.g. as Cu-HighGain-
HighResolution.bfcfg and Cu-HighGain-HighResolution.zip from the CONFIGURATION and DETEC-
TOR plugin.
In the same way the detector can be energy calibrated for High Count Rate mode. The spatial calibra-
tion remains the same and needs not to be re-determined.