EasyManua.ls Logo

BSWA 308 - Page 37

BSWA 308
126 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
BSWA 308/309 BSWA Technology Co., Ltd
35
(3) Select Calibration in the menu and then press <Enter> to enter By Measurement.
(4) Adjust Cal.Level in the menu, for example adjust to 93.8dB. There is 5s delay after
press <Start> to run calibration.
(5) After the end of calibration, sound level meter will update the calibration factor. Press
<Enter> to apply the results.
(6) Return to Main screen and press <Start/Stop> to start the measurement. The current
measurement result will be 93.8dB in this example if calibrator is still working.

Table of Contents

Related product manuals