BSWA 308/309 BSWA Technology Co., Ltd
72
Example: set to enable response.
02 01 43 52 45 54 31 03 31 0D 0A
Return: ACK.
RET?: Query Response Mode Setting
Example: query response mode.
02 01 43 52 45 54 3F 03 3F 0D 0A
Return: the current response mode is to enable the response.
MEMp1: Set the Measurement Mode
When MEM instruction is correctly received by the sound level meter, it will switch to the main
screen of the octave mode or the main screen of level meter mode according to the
corresponding parameter in instruction.
★Note: The 1/3 octave band is optional function.
p1: Measurement mode;
0=1/1Octave;
1=Level meter mode;
2=1/3 Octave (Optional);
Default: 1