Chlorine sensor
568524
1)
and 568523
2)
Instructions "Trace" chlorine sensor
565164
—
Versions with the internal holder:
1. Carefully fill up the internal holder with the electrolyte.
2. Make sure there is no air (bubbles) in the internal holder, else do the
filling again.
—
3. Hold the shaft upright and push the working electrode carefully into the
filled internal holder.
—
4. Check if air bubbles are visible in the specific membrane of the internal
holder.
5. If there are air bubbles, remove the internal holder from the shaft if
needed, rinse it carefully with clean water and dry the specific mem-
brane with an absorbent paper towel.
6. Repeat instructions 1 to 5 until there are no more air bubbles in the
internal holder.
16
Installation and wiring
Type 8232
english