CX3001 v7.0 © Calnex Solutions 92
problem does not clear contact
Calnex support.
Battery charge time excessive,
Battery Fault
Battery has failed to charge within the
specified time limit, indicating a
battery fault.
Indicated by battery status
icon. Export a problem report
and shutdown Sentinel. Contact
Calnex support.
Battery has reached full capacity,
charging is disabled.
Temperature sensors reading
failed
Unable to read from processor board
temperature sensors
Check cooling fan on side of
Sentinel is spinning. If fan is not
spinning or the problem does
not clear, export a problem
report and power cycle Sentinel.
Temperature sensor readings
restored
Successful read from processor board
temperature sensors after previous
failure
None. Sensor communication
error has cleared.
Enable calibration options
A problem report can be generated by pressing the Export problem report… button on the
About page of the System menu and saving to internal memory or an external USB drive. If after power cycling
Sentinel, the problem does not clear then the problem report should be forwarded to Calnex along with a
description of the fault and a block diagram of the test setup.