EasyManua.ls Logo

Cameron BARTON 289A - Calibration Check

Cameron BARTON 289A
75 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
15
Model 289A Differential Pressure Indicating Switch Section 3
Bezel
(9A-C0277-0029C)
Cover (Lens)
(9A-C0181-0038C)
Gasket
(9A-C0277-0026C)
Case
Figure 3.1—Bezel and lens cover
IMPORTANT: Ensure the bezel gasket is properly oriented before placing the instru-
ment back in service. Incorrect gasket orientation will cause the instru-
ment indicator to jam, resulting in inaccurate readings.
Calibration Check
To ensure the unit calibration is within factory-set calibration tolerances,
perform the following procedure.
IMPORTANT: Review all procedures, WARNINGS/NOTICES in the Model 199 DPU
user manual (Appendix A) BEFORE performing this procedure.
HUB
Pointer "slipped"
rotated on hub
Hub held in place
with 1/4" wrench
Figure 3.2—"Slipping" pointer
1. Mount the instrument in an approximately level position and connect to
a standard pressure source (see the Model 199 DPU user manual in Ap-
pendix A).
2. If the zero indication is incorrect, adjust it as follows:
a. Remove the bezel/lens assembly.
b. Using a 1/4" open-end wrench (included in calibration toolkit, Part
No. 9A-0288-1032B), hold the hexagon pointer hub xed and rotate
the pointer with ngers until the pointer indicates zero on the scale.
See Figure 3.2.
c. Replace the bezel/lens assembly.

Table of Contents

Other manuals for Cameron BARTON 289A

Related product manuals