INSTRUCTIONS FOR USE
CapsoAccess
®
Capsule Data Access System
Catalog Number: CDAS3
English Nederlands Français Deutsch Italiano Español

Table of Contents

Questions and Answers

  • S
    Sarah GardnerAug 5, 2025
    How to enable 'Download Images' in CapsoView® Medical Equipment?
    • K
      Katherine LeblancAug 5, 2025
      If the 'Download Images' option is grayed out in CapsoView®, it means it's not enabled. Here’s what you can do: 1. Check the USB cable connection between the system and your computer. Ensure the system power is ON and the indicator light is lit. 2. Make sure your computer is set up to download capsule data and that the system USB driver is installed. 3. Try turning the system OFF, waiting a few seconds, and then turning it back ON. 4. Disconnect and then reconnect the USB cable. 5. Restart the CapsoView® software. 6. Reboot your computer.
  • T
    Teresa DavisAug 12, 2025
    What to do if a system data transfer error message appears after selecting 'Download Images' on CapsoVision Medical Equipment?
    • D
      Dennis RogersAug 12, 2025
      If you get a system data transfer error after selecting “Download Images,” try these steps: 1. Turn the system OFF, wait a few seconds, and turn it back ON to reset the USB connection. Then, try downloading again. 2. Try using a capsule that you know works well (one with previously downloaded data). If the error message appears again, the system itself might be malfunctioning. 3. If the known-good capsule docks successfully and the Procedure Information window appears, the original capsule may be damaged.
  • F
    flesterAug 19, 2025
    What to do if CapsoVision Medical Equipment says no capsule is detected after selecting 'Download Images'?
    • J
      jose08Sep 12, 2025
      If you receive an error message stating that no capsule is detected after selecting “Download Images”, consider the following: * Make sure the capsule is correctly placed in the receptacle, with the battery-end facing upwards. * Check that both the capsule dimple and the system receptacle are clean. * Try rotating the capsule by approximately 180° and then try again.
  • L
    Laura LoveAug 27, 2025
    Why is the CapsoAccess® Medical Equipment system indicator light not on?
    • H
      Heather SnyderAug 27, 2025
      If the CapsoAccess® system indicator light isn't lit, check these: * Ensure the power switch is in the ON position. * Verify the power cable is fully plugged into the outlet and the power supply, and confirm the light on the power supply is illuminated.

Summary

Safety Precautions

Electrical Shock and AC Adapter Safety

Specific safety caution regarding electric shock from the AC adapter and proper connection.

Getting Started

Principle of Operation

CapsoAccess System Operation

Explains the inductive power, optical link, and data transfer mechanism of the system.

Installation Guide

System Operation

Troubleshooting Guide

Care, Maintenance, and Disposal

Cleaning Procedures

Provides detailed instructions for cleaning the system's exterior and receptacle.

Disposal Guidelines

Advises on the environmentally sound disposal of the system and its power supply.

Specifications and Ratings

CapsoVision CapsoAccess CDAS3 Specifications

General IconGeneral
BrandCapsoVision
ModelCapsoAccess CDAS3
CategoryMedical Equipment
LanguageEnglish