8400-0181-OM Rev C 185 and 185B Installation & Technical
20
BR= (Break Range)
The break range is difference in divisions between consecutive samples, which will turn
filtering off. Filtering will resume when the difference between samples is less than the break
range. A smaller value for break range will result in a display that is more responsive to
changes in weight.
Press the TARE key to view the current setting. To keep the current setting, press the
TARE key again. Otherwise, use the Fn/ and UNITS/ keys to enter a new value.
When the desired value is displayed, the TARE key to select the displayed value and
proceed to the next prompt.
Available values are 0 (disables break range) through 99.
CALgC= (Calibration Location Gravity Constant)
This is the acceleration of gravity value of the location where the scale was calibrated.
NOTE: Set to zero (0) if gravity constants are not going to be used.
Press the TARE key to view the current setting. To keep the current setting, press the
TARE key again. Otherwise, use the Fn/ and UNITS/ keys to enter a new value. When
the desired value is displayed, the TARE key to select the displayed value and proceed to
the next prompt.
Consult the factory Tech Support for the appropriate value to enter. Note that if a nonzero
value is entered, then an additional prompt will appear.
oPgC= (Operating Gravity Constant)
This is the acceleration of gravity value for the location where the scale will be operated.
Press the TARE key to view the current setting. To keep the current setting, press the
TARE key again. Otherwise, use the Fn/ and UNITS/ keys to enter a new value. When
the desired value is displayed, the TARE key to select the displayed value and proceed to
the next prompt.
Consult the factory Tech Support for the Acceleration of Gravity value for your location.
PUO = (Power-Up Zero Feature)
Press the TARE key to view the current setting. To keep the current setting, press the
TARE key again. Otherwise, press the Fn/ key to scroll through the available values.
When the desired value is displayed, press the TARE key to save it.
Available values are:
YES no
utomatic Re-Zero on Powe
r
-Up No Re-Zero on Power-Up
ASH= (Automatic Shutoff)
Automatic Shutoff will automatically turn the indicator off (when it is not in use) after a period
of inactivity. To turn the indicator back on you must press the ON/OFF key.
Press the TARE key to view the current setting. To keep the current setting, press the
TARE key again. Otherwise, press the Fn/ key to scroll through the available values.
When the desired value is displayed, press the TARE key to select the displayed value
and proceed to the next prompt.
Available values are 0 (automatic shutoff disabled) through 9.