8555-M210-O1 Rev R 758C Owner’s 21
SETUP AND CALIBRATION, CONT.
CAP= (SCALE CAPACITY)
With display showing CAP=, press PRINT/ENTER key to show current setting. If value
displayed is acceptable, press PRINT/ENTER key to save it. Otherwise, using numeric keys,
enter a new value and then press PRINT/ENTER key to save it. Allowable values are 1
through 99,999.
CAL= (CALIBRATION)
With display showing CAL=, press PRINT/ENTER key. Display will change to show current
setting 0 (0=NO). If scale has been previously calibrated and you wish to skip calibration and
proceed to trA=, Zero Tracking Range, simply press PRINT/ENTER key and internal
calibration factor will be retained.
To begin calibration, press 1/YES key to select yes and then press PRINT/ENTER key. After
pressing PRINT/ENTER key display will change to LOAd=.
LOAd= (LOAD CALIBRATION WEIGHT)
Display will change to LOAd= which is a prompt for entry of calibration weight value and
placement of this amount of test weights on scale platform.
1. Make certain scale platform is empty and free of debris.
2. Place the desired amount of calibrated test weights on scale platform. A minimum of
50% of scale's capacity is required. However 70% to 100% is recommended.
3. Press PRINT/ENTER key.
4. If the value displayed is acceptable, press the PRINT/ENTER key again. Otherwise,
determine the exact amount of test weights to be placed on scale platform, then using
numeric keys, enter this value.
5. Verify that the numbers entered are the same as the amount of test weights and then
press the PRINT/ENTER key.
6. Starting at the left and proceeding right, a series of dashes will appear on the display.
The dashes will stay on the display momentarily, then disappear, after which the display
will proceed to the UnLOd prompt.
UnLOd= (UNLOAD CALIBRATION WEIGHT)
With display showing the UnLOd prompt:
1. Remove the weights from the scale platform and then press the PRINT/ENTER key.
2. Starting at the left and preceding right, a series of dashes will appear on the display.
The dashes will stay on the display momentarily, then disappear, after which the
calibration factor will be saved and the display will proceed to the trA= prompt.
trL= (FOUR PERCENT ZERO TRACKING RANGE LIMIT)
This setting determines whether a 4% limit is placed on zero tracking and push button zero
and selects the format when printing the date.
If enabled, (trL= 1) zero tracking and push button zero is limited to 4 % of scale capacity
and the date format is dd/mm/yy.
If disabled, (trL= 0) zero tracking and push button zero will operate up to full scale
capacity and the date format is mm/dd/yy.
With display showing trL=, press PRINT/ENTER key to show current setting. If setting
displayed is acceptable, press the PRINT/ENTER key to save it. Otherwise use numeric
keys 0/NO or 1/YES and enter a new setting. Press PRINT/ENTER key to save it.

Questions and Answers: