Page | 28
• Last calibration date
• Display board firmware version
• CDU board firmware version
• Check on battery voltage
• Battery type selected (default lithium-ion or user-selected lead acid)
• Checks on USB formatting and available memory
• Checks on motor controller
If the USB drive is not present, an error appears on the splash screen showing ‘USB Setup Fault’. Turn off the
instrument by disconnecting the power source, insert the USB drive and power up again.
If the USB drive has not been formatted correctly, an error message appears as shown in Figure 24.
In this case, press ENTER to reformat the USB drive or press 1 to abort the boot-up sequence. If the reformat option
is selected, this process will take around one minute, after which the instrument will turn itself off. All data on the
USB drive will be deleted when it is reformatted.
Once the start-up checks are complete, the instrument searches for existing FSC files on the USB drive. Only files
named QM_nnn.FSC will be recognised. If no such files are found, the unit displays a message stating that a new
file will be created: QM_001.FSC. Press ENTER to continue.
If any valid FSC files are detected, the unit identifies the one with the highest sequential number. An option is then
given to either append to this file, or to start a new file. In the case shown in Figure 26, file QM_001.FSC was found.
Figure 25 Data not found
Figure 24 USB drive error and reformatting