110 
 
10.6 Check Sensor Accuracy  
Place the sensor in a medium of known temperature and compare that temperature to the measured reading. The 
thermometer  used  to  determine  the  temperature  of  the  medium  should  be  of  laboratory  quality  with  0.25  °C 
graduations. The sensor in question should be accurate to within 1.2 °C. 
 
Note that the PIC5+ control module, MAINTENANCE menu, offers a temperature sensor calibration feature where 
the sensor temperature can be offset. Note that only the four water temperatures can be calibrated. To use this 
feature,  place  the sensor @  0°C  or  other  known  temperature.  Read  the  raw  temperature  and  calculate  offset 
based on the reading seen in the TEMP_CAL menu. Enter and execute the offset, which cannot exceed ± 1.2°C. 
 
The sensors are immersed directly in the refrigerant or water circuits. When installing a new sensor, apply a pipe 
sealant or thread sealant to the sensor threads. 
An additional thermistor, factory installed in the bottom of the cooler barrel, is displayed as Evap Refrig Liquid 
Temp on the TEMPERATURES display screen. This thermistor provides additional protection against a loss of 
water flow. 
 
DUAL TEMPERATURE SENSORS  
For servicing convenience, there are 2 redundant sensors each on the bearing and motor temperature sensors. If 
one of the sensors is damaged, the other can be used by simply moving a wire. The number 2 terminal in the 
sensor terminal box is the common line. To use the second sensor, move the wire from the number 1 position to 
the number 3 position. See fig 59 and 60. 
 
 
 
Fig 59 – 1
st
 stage / 2
nd
 stage Bearing Temperature (RT39C / RT40C) sensor wiring 
 
 
 
 
Fig 60 – Winding Temperature (RT31C) sensor wiring