27
1. General Function
Menu-GEN.01: Set AD Speed
Function AD Speed Setting
Display Part Meaning
ADC. 0 Converting speed 5 times per second
ADC. 1
Converting speed 10 times per second
ADC. 2 Converting speed 20 times per second
ADC. 3 Converting speed 40 times per second
ADC. 4 Converting speed 100 times per second
ADC. 5 Converting speed 160 times per second
Set Range
(0 ~ 6)
ADC. 6 Converting speed 320 times per second
Note 1. After changing the AD speed, press āSet' key to display the weight using the altered value.
(Press ā
Set' after checking the changed value to return to the previous mode)
Menu-GEN.02: Set average filter
Function Average filter Setting
Display Part Meaning
Set Range
(1 ~ 50)
Fil. XX
Initial Value : 10
Setting the number of buffers in the average filter
Note 1. Set based on the environment. (The rate of change of weight may slow)
Menu-GEN.03: Set digital filter
Function Average filter Setting
Display Part Meaning
Set Range
(0 ~ 5)
IIR. XX
Initial Value : 0
0 : Filter not use
1 ~ 5 : Filter intensity
Note 1. Set based on the environment. (The rate of change of weight may slow)
Menu-GEN.04: Set use key lock
Function Set use key lock
Display Part Meaning
LOCK. 0
Unlock front key
Set Range
(0 ~ 1)
LOCK. 1 Lock front key
Menu-GEN.05: Set Key Operating Conditions
Function Key Operating Conditions(Zero, Tare) Setting
Display Part Meaning
K-ST. 0 Always in operation
Set Range
(0, 1)
K-ST. 1
Operates only if the weight is stable