CAL 6
Function: Zero adjustment - calibration when any zeroing error occurs.
ZERO Key : Cancel
SET Key : Save and next
menu
Empty the load tray and press the setup key.
The current weight value is displayed.
Confirm 'Stable' and press the setup key.
Zero adjustment in progress...
Note 1. Use this function when zeroing is not passed for any shock to the load cell.
The range of zero adjustment is 0 ~ 2mV/V.
CAL 7
Function: Factor Calibration
▲ Key : Value change
◀ Key : Digit change
ZERO Key : Cancel
SET Key : Save and next
menu
This function cannot be used because of multi
calibration.
It means you entered the factor correction mode.
The current factor is displayed.
Note 1. As this is a menu to set the weight setup with no weight, general users have no need to
use it.
Note 2. This can be used only when the range of multi calibration in CAL 3-1 is set to 1.
"NOT USE" is displayed when the range of CAL 3-1 is set to 2 or larger.
Note 3. Enter a password to enter the factor correction mode.