Menu-GEN.09: Set Zero Key Range
Function
Zero Key Range Setting
Set Range
(1 ~ 99)
Display Part
Meaning
ZKR. XX
Initial Value :
10 %
Zero operates up to within +/- 慇慇% of the maximum weight
Menu-GEN.10: Set Tare Key Range
Function
Tare Key Range Setting
Set Range
(1 ~ 100)
Display Part
Meaning
TKR. XX
Initial Value :
100 %
Tare operates up to within +/-慇慇% of the maximum weight
Menu-GEN.11: Set Initial Zero Range
Function
Initial Zero Range Setting
Set Range
(1 ~ 99)
Display Part
Meaning
ZIR. XX
Initial Value :
10%
Initial zero operates within +/-慇慇% of the Gross Weight
Menu-GEN.12: Set Overload Range
Function
Overload Range Setting
Set Range
(1 ~ 99)
Display Part
Meaning
OVRD. XX
Initial Value :
9 x 1 Digit
Overweight from the next to 0 x 1 Digit of the maximum weight
Menu-GEN.13: Set Hold Type
Function
Hold Type Setting
Set Range
(0 ~ 3)
Display Part
Meaning
HOLD.00
Average Value Hold
HOLD.01
Peak Hold
HOLD.02
Sampling Hold
HOLD.03
Automatic Hold
Menu-GEN.14: Set Hold Canceling Conditions
Function
Hold Canceling Conditions Setting
Set Range
(0 ~ 2)
Display Part
Meaning
HD-C.00
Hold is canceled when it becomes zero
HD-C.01
Hold is canceled when Hold key is entered
HD-C.02
Hold is canceled when Hold less key entered