Register List
The Modbus-RTU protocol registers capable of being executed in this instrument are as
follows.
R = Register that can be used as read-only
W= Register that can be used as write-only
R/W = Register that can be used as for both read and write
H = Higher word of the Double word composing the register
L = Lower word of the Double word composing the register
REGISTER
DESCRIPTION
Input value
Saving to
EEPROM
ACCESS
40002
Type of instrument
-
-
R
40008
GROSS WEIGHT H
-
-
R
40009
GROSS WEIGHT L
-
-
R
40010
NET WEIGHT H
-
-
R
40011
NET WEIGHT L
-
-
R
40014
Raw AD Data_H
-
-
R
40015
Raw AD Data_L
-
-
R
40017
Set point 1 H
0~99999
Y
R/W
40018
Set point 1 L
0~99999
Y
R/W
40019
Set point 2 H
0~99999
Y
R/W
40020
Set point 2 L
0~99999
Y
R/W
40021
Set point 3 H
0~99999
Y
R/W
40022
Set point 3 L
0~99999
Y
R/W
40023
Set point 4 H
0~99999
Y
R/W
40024
Set point 4 L
0~99999
Y
R/W
40037
Ext_Input
-
-
R/W