③ Manual accumulation of measured value:
Press the [ACCUM] key when the measured value is larger or equal to 20 divisions and the data is
stable in the normal weighing state, the indicator will perform "manual accumulation" function. In
this case, the indicator will display the total accumulation data (in two steps): [total = ] (indicating
that the content shown below is the amount of accumulation data) will show the accumulation
data [******] in about 1 second. The times of accumulation will be then indicated (in two steps):
[n = ] (indicating that the times is shown below) will show the times of accumulation [ ***] in
about 1 second. The ACCUMULATION indicating light is then on.
Note : the maximum times of accumulation is 9999 (when accumulation result is ensured to be
≤999999); The accumulation results will be held before they are cleared and the data will not be lost
after power-off. When the [ACCUM] key is pressed, if the net weight is less than 20 divisions, only the
accumulation value will be displayed, and the accumulation of weight and times will not be made.
④ Automatic accumulation of measured value:
Press the [FUNC] and [ACCUM] key at the same time in normal weighing state and not the peak
value holding and average value state, the indicator will enter the automatic accumulation state. In
this case, the ACCUMULATION light is on. In the automatic accumulation state, the indicator will
perform an automatic accumulation and display the times of accumulation and the results
whenever the measured data is larger than or equal to 20 divisions and after about 1-2 seconds after
the data is stable. Re-press the [FUNC] and [ACCUM] keys at the same time or press [HOLD]key
to exit the automatic accumulation.
Note : automatic accumulation state will not be saved when the power is off; however, the
accumulation data will be saved.
⑤ Clear the accumulation result :
Press the CLEAR key in weighing state, the indicator will clear the times of accumulation and the
overall accumulation data. The user should Clear first and then do the operation in first time
accumulation.
⑥ Switching between kg and lb :
Press the [FUNC] and [HOLD] keys at the same time in normal weighing state, the indicator will
perform switching between units; when the measurement unit is lb, the last digit behind decimal
point of the shown data is on. The users can also switching unit by adjust the parameter of P1.
⑦ Preset tare and upper & lower limits alarm:
Press the [HOLD] keys for long in normal weighing status in the following steps:
Weighing indication status
Press [FUNC]
For long time
Guide the user to enter preset tare value. Preset tare value can
be set within 5 numbers value.
Input the preset tare
value, e.g. “6000”
Press the [HOLD] key to make confirmation and move to step 4.
Upper limit alarm
value, e.g. “3000”
Press the [HOLD] key to make confirmation and move to step 5.
Lower limit alarm
value, e.g. “0”
Press the [HOLD] key to make confirmation and move to step 6.
Go back to weighing mode. Displaying net weight after tare.
Note : Cancel the upper and lower limit alarm is not workable when both the upper and the
lower limit is 0;