EasyManua.ls Logo

CAS DH - Calibration Procedures; Entering Calibration and Division Setting; Decimal Point and Full Capacity Setting

CAS DH
19 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
11
Calibration Description
Properly connect the signal source and power supply to preheat the indicator for 15-30
minutes when there is no load on weighing platform.
After lead sealing is broken, please insert short connecting ring to connect pin J7 on the indicator
main board, and allows the indicator to be calibrated.
circuit board is setting as default that allows calibration, new indicator can omit this step),then
operate as following stepsafter calibration, lead sealing again):
(1) Press [HOLD] key during startup initialization, the indicator will come into calibration state.
Division setting
Display d X press [TARE] key to choose 1,2,5,10,20,50press [HOLD] key for
confirmation, and will enter into next parameter setting,
press [TARE] key for automatic step-by -step cycle display.
Display d X
Display d 1
Display d 2
Display d 5
Display d 10
Display d 20
Display d 50
Display d 1
For example, press [HOLD] key when it showsd 5,now the division is setting to 5, and
will come into decimal point setting state automatically.
decimal point setting
Display P X press [TARE] key to choose 0123press [HOLD] for
confirmation, and will come into next parameter setting automatically. Press [TARE] key for
automatic step-by-step cycle display.
Display P 0
Display P 0.0
Display P 0.00
Display P 0.000
Display P 0
For example, press [HOLD] key when it showsP 0.000,now the decimal point setting is
0.000and will come into full capacity setting state automatically.

Related product manuals