32
F05
FUNCTION : Stable condition of weight
VF DISPLAY SUB DISPLAY DESCRIPTION
23
Stable lamp is off even with the
change of only two division for
3sec.
55
Stable lamp is on if the weight is
changed within five division for 5
sec.
Set value
(00~99)
98
F05 StAbLE
Stable lamp is on if the weight is
changed within nine division for 8
sec.
Note 1. The first number indicates division and the second number indicates second on the
VF Display.
F06
FUNCTION : Automatic zero condition
VF DISPLAY SUB DISPLAY DESCRIPTION
00
No compensation
23
Compensation for gradual change
below 1 division for 3 sec.
Set value
(00~99)
99
F06 AZEro
Compensation for gradual change
below 4.5 division for 9 sec.
Note 1. To have division, divide the first number on the VF Display by 2.
Note 2. The second number indicates a second on the VF Display..
F07
FUNCTION : Weight backup
VF DISPLAY SUB DISPLAY DESCRIPTION
oFF Weight backup is off
Set value
(oFF~on)
on
F07 bACKUP
Weight backup is on
Note 1. If weight backup is on, the scale saves previous weight when power failure is
occurred.
Note 2. On and Off are alternately displayed by pressing the
key.