Questions and Answers

  • K
    Kevin CameronAug 19, 2025
    What to do if CAS Accessories show Err 02?
    • W
      William RodriguezAug 19, 2025
      If your CAS Accessories display 'Err 02', it indicates a load cell connection failure or an error in the A/D conversion. To resolve this, check the L/C connector and test the A/D conversion in Test Mode 2.
  • B
    Brian Salas PhDAug 21, 2025
    What to do if CAS Accessories show Err 01?
    • A
      Aaron WhiteAug 21, 2025
      If your CAS Accessories display 'Err 01', it means the initial value of A/D has failed. You should check the L/C connector and test the A/D conversion in Test Mode 2 to fix this.
  • J
    joshuacharlesAug 23, 2025
    What to do if CAS Accessories show Err 13?
    • D
      Diana MillerAug 23, 2025
      If CAS Accessories show 'Err 13', it indicates that the zero value exceeds the initial zero range. To fix this, remove some weight from the platter and then turn the power on.
  • M
    Michael Stewart Jr.Aug 26, 2025
    What to do if CAS Accessories show Err 05?
    • H
      Henry FernandezAug 26, 2025
      If your CAS Accessories display 'Err 05', it means the input value is over the acceptable range. Input a value between 32 and 255 to resolve this issue.
  • N
    nathanielcruzAug 27, 2025
    What to do if CAS Accessories show Err 08?
    • K
      Kim DeleonAug 28, 2025
      If CAS Accessories show 'Err 08', the Hi limit weight is larger than the Maximum weight. Set the Hi limit weight to under the maximum weight.
  • J
    juanleonardAug 30, 2025
    What to do if CAS Accessories show Err 09?
    • J
      John BatesAug 30, 2025
      If CAS Accessories display 'Err 09', it means the Low limit weight is larger than the Hi limit weight. To correct this, set the Low limit weight to under the Hi limit weight.
  • F
    figueroajonSep 1, 2025
    What to do if CAS Accessories show Err 22?
    • D
      David Hernandez MDSep 1, 2025
      If CAS Accessories show 'Err 22', it means the weight for span calibration is set to under 10% of maximum capacity. The weight of Span calibration Should be set to over 10% of max. Capacity (Check it in CAL 1).
  • G
    gregorysnyderSep 3, 2025
    What to do if CAS Accessories show Err 23?
    • R
      Riley RichardsonSep 3, 2025
      If CAS Accessories show 'Err 23', it means the weight for span calibration is set to over 100% of maximum capacity. The weight of Span calibration Should be set to under 100% of max. capacity (Check it in CAL 1).
  • S
    Sara KochSep 5, 2025
    What to do if CAS Accessories show Err 21?
    • C
      Christian BrownSep 5, 2025
      If CAS Accessories display 'Err 21', it indicates that the resolution exceeds 1/20,000. To resolve this, lower the resolution.
  • M
    Michael MurphySep 7, 2025
    What to do if CAS Accessories show Err 07?
    • J
      John MillerSep 7, 2025
      If CAS Accessories display 'Err 07', it means the Product ID is larger than 19. The range of Product ID is 0 to 19.

Summary

Precautionary Measures

Installation Guide

Load Cell Connection Procedure

Details how to connect the load cell to the indicator.

Test Mode Operations

System Mode Configuration

System Mode Operations Overview

Outlines the different operational modes available within System Mode.

Count Mode Details

Provides detailed instructions and settings for the Count Mode.

Percent Mode Details

Explains the settings and procedures for the Percent Mode.

Accumulation Mode Details

Covers the operations and settings for the Accumulation Mode.

Calibration Mode Guide

Weight Calibration Steps

Details the steps for performing weight calibration.

Capacity and Division Settings

Setting maximum capacity and minimum division for calibration.

Zero Calibration Procedure

Details the zero calibration steps using the PRINT key.

Span Calibration Procedure

Details the span calibration steps.

Function Mode Settings

Using the Weighing Indicator

Communication Protocols

PC Connection via COM1

Details on connecting the indicator to a PC via COM1.

Data Protocols and Commands

CAS Protocol Details

Explanation of the CAS protocol structure and data format.

Limit Protocol Details

Explanation of the Limit protocol structure and data format.

Command Mode Protocol

Details the commands for controlling the indicator.

Error Messages and Troubleshooting

Error Code Explanations and Solutions

Lists error codes, their reasons, and solutions for troubleshooting.

CAS WEIGHING SOLUTION NT-200 Series Specifications

General IconGeneral
ModelNT-200 Series
CategoryAccessories
DisplayLCD
BrandCAS
Weighing Unitsg, oz