23
* Weight Calibration :: ( ZERO KEY - )
UNIT : 0 : kg 1 : lb
CAL 0 : Multi Calibration Setting
CAL 1 : Maximum Capacity Setting
CAL 2 : Minimum Division Setting
CAL 3 : Zero Calibration
CAL 4/6/8/10 : A Weight Setting (First~Fourth)
CAL 5/7/9/11 : Span Calibration (First~Fourth)
SELECT the UNIT of Weight (USA Version)
KEY FUNCTION DISPLAY VAL. DESCRIPTION
Select 0 or 1
ENTER
UNIT
Save and
Go to next
Unit -
0
1
kg
lb
CAL 0
MULTI - CALIBRATION SETTING
KEY FUNCTION DISPLAY DESCRIPTION
CAL 0
CAL 0 Condition
P = 1
1 Point Multi-Calibration
P = 2
2 Point Multi-Calibration
P = 3
3 Point Multi-Calibration
Increase or
decrease
set value
P = 4
4 Point Multi-Calibration
ENTER
UNIT
Next Menu Save the value and go to CAL 1
1 (OFF) : You can calibrate indicator with zero calibration and span calibration.
2 ~ 4 (ON) : You can set the weight what you want in the zero calibration and
the setting weight (CAL 4,6,8,10) and span calibration (CAL 5,7,9,11)