25
CAL 3
ZERO CALIBRATION
KEY FUNCTION DISPLAY DESCRIPTION
CAL 3
CAL 3 Condition
ULoAd
Remove weight from the platter
and Press UNIT KEY
---------
Good
Performing Zero Calibration
Zero Calibration is finished
ENTER Next Menu
Save the value and go to CAL 4
Note 1. If zero calibration is finished successfully, the display shows “Good" and
then you will go to CAL 4 automatically.
CAL 4
WEIGHT SETTING
Available Setting Range : 1kg ~ 999,999kg
KEY FUNCTION DISPLAY DESCRIPTION
CAL 4
CAL 4 Condition
Increase or decrease
set value
1000
1000kg
Move one digit to left
2000
2000kg
ENTER Next Menu
Save the value and go to CAL 5
Note 1. If you do not use multi calibration (P = 1),
the range of setting weight should be from 10% to 100% of maximum capacity.
Note 2. If you use multi calibration (P >=2),
the range of setting weight should be from 10% to 100% of maximum capacity in
CAL 4, CAL 6, CAL 8 and CAL 10.
Note 3. When you set the value of weight, it means not the weight on the plat- Form + the
weight being raised on but the weight being raised on the platform