26
CAL 5
SPAN CALIBRATION
KEY FUNCTION DISPLAY DESCRIPTION
CAL 5
Put the weight that is set in CAL 4
and then press the UNIT KEY
LoAd1
Repeat this step according to multi
calibration setting
---------
Good
Save
Performing Span Calibration
Span calibration is finished
Remove the weight,
press ENTER KEY Move to the
initial CALMOd
Note 1. If span calibration is finished successfully, the display shows “Good"
Note 2. If span value is low, the display shows error message (Err 24).
In this case, lower the resolution
※ Repeat CAL 4 and CAL 5 according to setting of CAL 0.
For example, if you set CAL 0 to 4, perform CAL 4, CAL 5, CAL 6 , CAL 7,
CAL 8, CAL 9, CAL10, and CAL11.
※ Example
Max. capacity: 5000 kg, Min. division: 1 kg, Multi calibration(CAL 0) : 4 point
A weight: 2000 kg 1EA, 1000 kg 2EA, 500 kg 1EA
CAL 0 : 4 CAL 1 : 5000 CAL 2 : 1 CAL 3 : Zero calibration
CAL 4 : 2000 CAL 5 : LOAD1 CAL 6 : 1000 CAL 7 : LOAD2
CAL 8 : 1000 CAL 9 : LOAD3 CAL 10 : 500 CAL 11 : LOAD4