15 | Page
INSTALLATION
POWER SOURCE REQUIREMENTS
When selecting a location for the unit, verify that each of the following requirements is satisfied:
Power Source: The wall power outlet must meet the power requirements listed on the unit data
plate. These units are intended for 110 – 120 VAC 50/60 Hz applications at the following
amperages:
TFO-1 TFO-3 TFO-5
• Wall power sources must be protective earth grounded and single phase.
• Wall power sources must conform to all national and local electrical codes.
• Supplied voltage must not vary more than 10% from the data plate rating. Damage to the
unit may result if supplied voltage varies more than 10%.
• The oven requires a dedicated 15-amp circuit.
• Use a separate circuit to prevent loss of product due to overloading or circuit failure. The
circuit must match or exceed the amperage requirement listed on the unit the data plate.
Power Cord: The unit must be positioned so that all end-users can quickly unplug the oven in the
event of an emergency.
• The unit comes provided with a 125 volt, 15Amp, 9ft 5 in (2.86m) NEMA 5-15P power cord.
• Always use this cord or an identical replacement.
Fuse: Each oven ships with a 16 amp 250V 5x20mm fuse installed in the power cord inlet.
• The fuse must be installed and intact for the unit to operate.
• Always find and fix the cause of a blown fuse prior to putting the unit back into operation.
NEMA 5-15R
wall socket