USER MAINTENANCE
Continued next page
Calibration continued
3
Place the oven in temperature calibration mode.
a. Press and hold both the UP and DOWN arrow
buttons simultaneously.
The Temperature Display will show the letters “C O”,
then begin flashing the current temperature value.
Note: If an arrow key is not pressed for five seconds, the
Temperature Display will cease flashing and store the last
displayed value as the new current chamber temperature
value.
Set Temperature
Set Temperature
Current Temp. Value
4
Adjust the current temperature value to match the
reference device.
a. Use the UP and DOWN arrow buttons.
Set Temperature
Corrected
Reference Device
5
After entering the correction adjustment, wait 5 seconds.
The temperature display will cease flashing
and store the correction as an offset.
The oven will now begin heating or passively
cool in order to reach your set point with the
corrected display value.
Set Temperature
Heating with Corrected
Value
6
Wait for the oven to achieve the new set point with the
corrected display value.