USER MAINTENANCE
End of procedure
Calibration continued
7
After the oven has achieved the set point,
wait for the oven chamber temperature to
stabilize.
The chamber temperature is stabilized
when no fluctuations of ±1°F or greater
have been detected with the reference
device for a minimum of 30 minutes.
Failure to wait for stabilization will result
in an inaccurate calibration.
Reference Device
8
Once the temperature has stabilized, compare the reference
device and the oven display temperature readings.
a. If the readings are the same, or the difference
between the two falls within the acceptable range of
your protocol, the oven is calibrated for
temperature.
-Or-
b. Advance to step 9.
Reference Device
Set Temperature
9
If the two readings still fall outside the acceptable range of
your protocol, repeat steps 38 up to two more times.
Three calibrations attempts may be required to
successfully calibrate ovens more than ±3F out of
calibration.
Reference Device
Set Temperature
If the temperature readings of the oven and the reference device fall outside your
protocol after three calibration attempts, contact Technical Support or your distributor
for assistance.