EasyManuals Logo

Casella CEL-350 dBadge User Manual

Default Icon
45 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #30 background imageLoading...
Page #30 background image
Casella CEL-35X dBadge Users Handbook – Page 30
Overload Indication: 140.4dB RMS (i.e. 0.1dB above the top of
the range),
Peak Overload Value: 143.4dB Peak Linear (i.e. 3.1dB above the
top of the range),
Stabilisation Time: 3 seconds after power up,
Operating Environment
Temperature: 0
o
C to +40
o
C,
Pressure: 65 - 108kPa,
Humidity ± 0.5dB: 30% - 90% (non-condensing),
Magnetic Fields: Negligible,
Storage Environment
Temperature: -10
o
C to +50
o
C,
Battery: Internal NiMH, 32 hour battery life,
Charge Time: Less than 90 minutes from discharged,
Dimensions mm (in): 72x47x52 (2.8x1.8x2.0),
Weight g (oz): 68 (2.4),
The CEL-35X/IS Intrinsically Safe dBadge is labelled
with the following:
ATEX:
I M1
II 1G
EEx ia IIC T4
EEx ia I
(T
a
= -20
o
C to +40
o
C)
7.2 Microphone Specification (CEL-252)
Classification: Type 2
Nominal Open-Circuit Sensitivity (250Hz): -28dB ± 3.0dB
(30mV/Pa)
Polarization voltage: 0V
Capacitance (polarized, 250Hz): 11-15pF
Operating temperature range: 0 to +40
o
C
Cartridge thermal noise: 25dB(A)
Dimensions: IEC 61094-4 Type
WS 2
FM/CSA:
Class 1
Division 1
Groups A, B C, D,
Temperature classification T4