EasyManuals Logo

CEM MARS 6 Synthesis User Manual

CEM MARS 6 Synthesis
80 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #55 background imageLoading...
Page #55 background image
51
Appendix A: GlassChem 20 Vessels
6. Place the turntable into the microwave system. Ensure the turntable is properly seated on the drive lug.
Load and perform a Method
1. View the “Fiber Optic Temperature Control” section for detailed instructions on connecting the ber optic probe
and entering the GF number.
2. Insert the ber optic probe through the open I/O port on top of the microwave cavity and into the control ves-
sel.
• The ber optic probe should slide easily into the thermowell through the frame screw and load disc until the
probe reaches the bottom of the thermowell.
• Do not try forcing the ber optic probe.
• If probe cannot be easily inserted, check vessel alignment. Loosen the screw and realign.
CAUTION
If resistance is felt when trying to insert probe, ensure that the thermowell is not broken inside the Teon
cover. The sharp edges of any break can restrict the movement of the probe.
3. Select/load the method from the Classic directory. See “Methods” section of this manual or contact CEM

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the CEM MARS 6 Synthesis and is the answer not in the manual?

CEM MARS 6 Synthesis Specifications

General IconGeneral
BrandCEM
ModelMARS 6 Synthesis
CategorySynthesizer
LanguageEnglish