67
28. Add the diluent to the sample in the
container on the external balance pan.
29. Wait for the weight to stabilize.
30. Press READY.
The instrument reads and records the weight of
the sample and diluent.
The instrument displays the weight of the
sample, the weight of the sample and diluent,
and the dilution ratio.
31. Press READY to continue the analysis.
32. Lift the cover of the SMART System
5
.
Place two glass fiber sample pads on the
balance pan. Close the instrument cover.
33. Press TARE.
Dilution (Ext)
Wt: 0.000g
Add diluent to container
and press READY.
ABORT
READING BALANCE . . .
Dilution (Ext)
Sample Wt: xx.xxxg
Sample + Dil Wt: xx.xxxg
Dilution Ratio: x.xxxx
Press READY to continue.
XXXXXX
Power: xxx %
Temp (xxx): xx.x
Time: xx:xx
Ratio: xx.xxxx
Wt: x.xxxxg
Place pads on balance and
press TARE.
ID RATIO DILUTION
1 2 3 4 5