EasyManuals Logo

CEM SMART System 5 User Manual

CEM SMART System 5
163 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #80 background imageLoading...
Page #80 background image
73
Note: The test screen appears indicating the
power, temperature, and time.
15. If Autotrack is off and a method identifi-
cation name or number is required, press
the operation key below ID.
16. Using the operation keys below the arrows,
position the cursor on (select) the first
number or letter of the sample identifi-
cation. Press ENTER. Continue to
position the cursor on each letter or number
and press ENTER until the identification
is complete.
Note: If using an identification number only, use
the numeric keys to enter the number.
17. Press READY to return to the test
screen.
18. Lift the cover of the SMART System
5
.
Place two glass fiber sample pads on the
balance pan. Close the instrument cover.
19. Press TARE.
20. Wait for the instrument to tare the weight
of the sample pads.
XXXXXX
Power: xxx %
Temp (xxx): xx.x
Dry Time: xx:xx
Wt: x.xxxxg
Place pads on balance and
press TARE.
EXTERNAL
ID SYRINGE
1 2 3 4 5
SAMPLE ID:
0 1 2 3 4 5 6 7 8 9 A B
C D E F G H I J K L M N
O P Q R S T U V W X Y Z
Space Delete
Highlight choice, press ENTER.
Press READY to continue.
XXXXXX
Power: xxx %
Temp(xxx): xx.x
Dry Time: xx:xx
Wt: x.xxxxg
Place pads on balance and
press TARE.
ID
1 2 3 4 5
TARING BALANCE . . .

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the CEM SMART System 5 and is the answer not in the manual?

CEM SMART System 5 Specifications

General IconGeneral
BrandCEM
ModelSMART System 5
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals