74
21. Lift the instrument cover. Remove the
sample pads from the balance pan. Place
the syringe containing the sample on the
balance pan. Close the instrument cover.
22. Press READY.
Note: Press the operation key below Abort to
end the analysis.
The instrument weighs the syringe containing the
sample.
Note: Press the operation key below Abort to
end the analysis.
23. Lift the instrument cover. Remove the
syringe from the balance pan.
24. Dispense the sample from the syringe on
the tared sample pads.
25. Place the empty syringe back on the
balance pan. Close the instrument cover.
26. Press READY.
The instrument reweighs the syringe.
Syringe Weigh (Int)
Remove sample pads. Place
filled syringe on balance
and press READY.
ABORT
WEIGHING SYRINGE . . .
Syringe Weigh (Int)
Dispense sample on pads.
Place syringe back on
balance and press READY.
ABORT
REWEIGHING SYRINGE . . .