58
Intelect
®
Mobile Ultrasound
MAINTENANCE
Maintaining the intelect Mobile UltrasoUnd
The following items should be checked at least monthly to ensure proper operation of this unit:
• Powercordandplug:Checktomakesurethecordisnotfrayed,kinked,anddoesnothavetornorcutinsulation.
• Applicatorcable:Checktomakesurethecableisflexible,freeofkinks,notfrayed,andtheinsulationisintact.
• Soundheadface:Checktoseethatthereisnobuild-upofgelorforeignmaterialonthealuminumface.
Cleaning
To clean the accessories, use only soap and water. Alcohol may be used to disinfect the aluminum surface, but avoid the plastic area.
The Intelect Mobile Ultrasound case may be cleaned by wiping with a damp cloth or mild cleaning solution. Avoid abrasive
cleansers.
NOTE: The sound head must be cleaned with alcohol between each therapy session.
Service
The Intelect Mobile Ultrasound must be recalibrated annually. It is recommended that all DJO ultrasound products be returned to
the factory or an authorized servicing dealer for repairs or recalibration. Recalibration is also recommended after the replacement or
repair of any major component. Should the unit require service, warranty, or repair, please contact the selling dealer or your local DJO
customer service.
EU Directive on Waste Electrical and Electronic Equipment (WEEE), ensures that product is appropriately disposed of or recycled at
the end of its life.