19
Intelect
®
Mobile Ultrasound
Sound Heads ............................................ 1 cm
2
, 2 cm
2
, 5 cm
2
, 10 cm
2
Duty Cycles
Pulsed ...........................................................................10%, 20%, and 50%
Continuous .............................................................................................100%
Pulse Frequency ....................................................... 16, 48, and 100 Hz
Pulse Duration ............................... 1 msec, +/-20% (10% Duty Cycle
.....................................................................................................Pulsed Mode)
............................................................... 2 msec, +/-20% (20% Duty Cycle
.....................................................................................................Pulsed Mode)
............................................................... 5 msec, +/-20% (50% Duty Cycle
.....................................................................................................Pulsed Mode)
Output accuracy .......................... +/- 20% above 10% of maximum
Amplitude ................................ 0 to 2.5 W/cm
2
in Continuous mode,
.........................................................................0-3 W/cm
2
in pulsed modes
Temporal Peak to Average Ratios:
2:1, +/- 20%, for 50% Duty Cycle
5:1, +/- 20%, for 20% Duty Cycle
9:1, +/- 20%, for 10% Duty Cycle
ULTRASOUND TECHNICAL SPECIFICATIONS
Maximum Treatment Time .................................................30 Minutes
Output
Pulsed ...........................................1 MHz signal,
modulated 100% by the
100 Hz rectangular wave
with the selected Duty Cycle.
Continuous .............................1 MHz, nominal signal
that is activated as long
as the timer is operating.
Timer Accuracy .................................................................... +/-0.2 minute
SPECIFICATIONS