5 Set-up and connection
5.1 Installation site
Operate the freeze-dryer only in closed and dry rooms.
NOTE
Refrigeration problems of the freeze-dryer are often caused by insufficient
conditions at the location of use. This is why compliance with the following
conditions is absolutely mandatory!
The table must be stable and have a solid, even tabletop.
Ensure sufficient ventilation. Do not place any paper, cloth, or similar
material behind or under the unit, since otherwise the air circulation will
be impaired.
Keep a safety distance of at least 30 cm around the freeze-dryer so that
the vents in the unit remain fully effective.
The ambient temperature must be in the range of +10°C to +25°C. A
potential night-time setback of the air conditioning system must be
taken into consideration.
Prevent the room temperature from rising, for example due to closed
doors at night.
Do not subject the freeze-dryer to thermal stress, e.g. by positioning it
near heat generators.
Prevent thermal overload, e.g. caused by other equipment in the direct
vicinity of the freeze-dryer.
Do not set up the vacuum pump directly next to the heat exchanger
(condenser).
In the case of water-cooled systems, ensure that the water circuit
provides a sufficient amount of cooling water.
Avoid direct sunlight (UV radiation).