Freeze-dryer Alpha 1-4 LSCplus
Freeze-dryer Alpha 2-4 LSCplus
7 Malfunctions and error correction
Version 01/2011, Rev. 2.3 of 03/07/2017 sb
83 / 131
Translation of the original operating manual
7.1.3 Insufficient ice condenser temperature
CAUTION
Ensure sufficient ventilation. Do not place any paper, cloth, or similar
material behind or under the unit, since otherwise the air circulation will be
impaired.
The refrigeration unit is equipped with a protective device against
overpressure in the refrigeration system and with a thermal motor
protection switch.
The protective devices trip
when the ambient temperature is too high
when the air circulation of the heat exchanger of the refrigeration
system is insufficient
when the refrigeration system is overloaded.
In these cases, the refrigeration unit will be switched off automatically. If the
permissible operating conditions are re-established after a cool-down
phase of several minutes, the refrigeration unit will be switched on again
automatically.
The malfunctions are displayed in the process and equipment information
window.
The minimum ice condenser temperature of approx. -55°C or approx.
85°C (depending on the type of freeze-dryer) is reached when the ice
condenser is not loaded and the ice condenser chamber is evacuated.
7.2 Process and error messages
The LSCplus control system displays the complete process and error
messages (see chapter 6.5.1.1 - "Main window "Manual"", dialog box
"Process and equipment messages"), which is why they are not included in
this operating manual.
You can order these documents from our service department.