EasyManua.ls Logo

Christ Beta 1-8 LSCbasic - Page 57

Christ Beta 1-8 LSCbasic
107 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Freeze-dryer Beta 1-8 LSCbasic
Freeze-dryer Beta 2-8 LSCbasic
6 Operation
Version 03/2018, Rev. 1.5 of 09/03/2020 sb-dc
57 / 107
Translation of the original operating manual
System properties (3)
This dialogue box is used to change the system settings.
Fig. 37: Dialogue box "System properties"
Beeper: The beeper sounds in the event of a malfunction, for example.
If the setting is "On", the beeper sounds at intervals of a few seconds
until the message is acknowledged.
If the setting is "Silent", the beeper sounds once when the malfunction
occurs.
If the setting is "Off", the beeper will not sound at all.
Click on touch: If this function is active, a clicking sound can be heard
whenever the system registers that a button has been touched.
Behaviour of the pressure control valve in the case of insufficient ice
condenser cooling: If this function is active (button "closed"), the pressure
control valve will close at an ice condenser temperature of 20°C during
the drying process in order to avoid damage to the vacuum pump caused
by the withdrawal of condensable gases. Selecting the button again
deactivates the function (button "controlled").

Table of Contents