7 Malfunctions and error correction
Version 03/2018, Rev. 1.5 of 09/03/2020 sb-dc
Translation of the original operating manual
7.1.2 Insufficient vacuum
NOTE
The vacuum checks must be carried out when the ice condenser is
frozen.
7.1.2.1 Small flange connections
Leakages are often due to improper small flange connections between the
various components and hose connections or to leakages in the valves.
Loosen the connection and place the centring ring (with sealing ring
inside) in a centred manner between the flange connections.
Seal the connection with the clamping ring by tightening the wing nut.
Ensure that the centring ring neither slips out of place nor gets jammed.
Fig. 52: Small flange and centring ring
Fig. 53: Small flange with centring ring
and small flange
Fig. 54: Attaching the clamping ring
Fig. 55: Tightened clamping ring