WarmAir
Model 135 Manual OPERATING INSTRUCTIONS
Page 13 of 43
Approvals
Electrical
EMC COMPATIBILITY TABLES ACCORDING TO IEC 60601-1-2
Emissions tests
Compliance
Electromagnetic environment guidance
RF emissions
CISPR 11
Group 1
The WarmAir
, Model 135 uses RF energy only for its internal
function. Therefore, its RF emissions are very low and are not likely
to cause any interference in nearby electronic equipment.
RF emissions
CISPR 11
Class A
The WarmAir
, Model 135 is suitable for use in all establishments
other than domestic and those directly connected to the public low-
voltage power supply network that supplies buildings used for
domestic purposes.
Fluctuations and Flicker emissions testing is not required or 120 or
100 Volt units
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Complies
16HV
MODEL 135, MEDICAL ELECTRICAL EQUIPMENT IN
ACCORDANCE WITH UL60601-1, IEC60601-1 AND
ASTM F2196-02.
ALSO CLASSIFIED WITH RESPECT TO ELECTRICAL
SHOCK, FIRE AND MECHANICAL HAZARDS ONLY IN
ACCORDANCE WITH CSA 22.2 NO. 601.1

Table of Contents