EasyManua.ls Logo

Cincinnati Sub-Zero WarmAir 135 - Inspection Form

Cincinnati Sub-Zero WarmAir 135
44 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
WarmAir
Model 135 Manual INSPECTION FORM
Page 24 of 43
Inspection Form
MAKE COPIES FOR PREVENTATIVE MAINTENANCE PROGRAM
WarmAir
System - Model 135 Unit
Functional & Safety Inspection and Preventative Maintenance
Location:
Performed by:
Date:
Unit Serial Number:
Hospital Control Number:
Every 6 mo.
1. Damage Inspection - Unplug Power Cord! Risk of Electrical Shock!
Pass Fail
X
1.1
External cabinet is free of damage and secure
X
1.2
Control panel is free of damage and legible
X
1.3
All labels are legible and secure
X
1.4
I.V. Pole Clamp and knob are functional and secure
X
1.5
Bed Hook is functional and secure
X
1.6
Power cord plug is free of damage and secure
X
1.7
Power cord and strain relief are free of damage and secure
X
1.8
Power switch is functional and secure
X
1.9
Hose is free of cuts or punctures
X
1.10
Rubber feet are free of damage and secure
X
1.11
Heater screen is free of damage, unobstructed, and secure
X
1.12
Air filter is free of damage, debris, and is secure
2. Functional Verification
Pass Fail
X
2.1
Plug unit into proper voltage outlet and turn the power switch to the
on position, all LEDs should blink twice and the audible beeper sounds
X
2.2
Press all four setting buttons sequentially, the corresponding LED and
audible signal activate
X
2.3
The blower activates when ‘FAN ONLY’ is selected
X
2.4
The blower and the heater activate when selecting a heat setting
3. Temperature Setting Verification
Pass Fail
X
3.1
Ensure hose is extended. Place temperature sensor in center of
airstream, flush with the hose fitting. (see Appendix 1 on page 43)
X
3.2
Turn unit on. Set unit on "Low". Run for 5 minutes.
Record temperature here. _____________
Acceptable temperature range defined in Appendix 1 on page 43.
X
3.3
Set unit on "Medium". Run for 5 minutes.
Record temperature here. _____________
Acceptable temperature range defined in Appendix 1 on page 43.
X
3.4
Set unit on "High". Run for 5 minutes.
Record temperature here. _____________
Acceptable temperature range defined in Appendix 1 on page 43.