WarmAir
Model 135 Manual TROUBLESHOOTING
Page 27 of 43
Troubleshooting
Call CSZ Medical Technical Support for Replacement Parts. If you wish to return the unit to CSZ for
examination and repair, please refer to the following section. In Europe contact our authorized
representative.
The Warming Unit does not
turn on.
Unplugged or damaged power
cord.
Make sure power cord is
plugged in and is undamaged.
Replace cord if necessary.
Check circuit breaker power
switch.
Reset power switch (turn power
switch to off position then back
to the on position).
Poor or loose wire connections.
Ensure all connectors and
terminals are secure.
Ensure power cord and the base
wire harness are properly
attached to terminal block.
Safety Thermostat(s) open.
Allow unit to cool from possible
over-heat condition. If condition
persists, check continuity of
each thermostat. Replace if
necessary.
The Warming Unit blows air
but does not heat.
Check resistance value of each
leg (White/Gray & Red/Green)
at room temperature.
18C/65F 3081
21C/70F 2690
24C/75F 2354
Replace sensor if necessary.
Verify heater header/terminals
are secure on the control board
(J4) as well as the main
connector (J10/P10).
Check the heater wires
resistance (approx. 12). If
open, replace heater.
Consult with authorized CSZ
representative or replace board.