WarmAir
Model 135 Manual TROUBLESHOOTING
Page 28 of 43
The Warming Unit heats up but
then shuts off before reaching
temperature.
Thermostat (heater/blower
housing) maybe out of
calibration.
Allow unit to cool from possible
over-heat condition. If
condition persists, check
continuity of each thermostat.
Replace if necessary.
Consult with authorized CSZ
representative or replace
board.
Check resistance value of each
leg (White/Gray & Red/Green)
at room temperature.
18C/65F 3081
21C/70F 2690
24C/75F 2354
Replace sensor if necessary.
Obstructed Filter or Blower
wheel.
Verify unit has adequate air
flow. Replace filter if necessary.
Remove top cover and
thermistor. Remove hose and
verify that blower wheel spins
freely.
The Warming Unit alarms at
too low or too high a
temperature.
Check resistance value of each
leg (White/Gray & Red/Green)
at room temperature.
18C/65F 3081
21C/70F 2690
24C/75F 2354
Replace sensor if necessary.
Confirm unit is being used in
ambient between 15°C/59°F
and 30°C/86°F. Consult with
authorized CSZ representative
or replace board.