Quality control recommendation
88
Applying the control solution
10 Apply a large free hanging drop of either the
glucose or cholesterol low level control
solution directly from the bottle to the test strip.
Ensure that neither the bottle nor your fingers
touch the application area.
The application area must be completely covered.
The sample can (as with performing a
measurement using blood) also be applied
outside the instrument as described on
page 68.